Agrochemical Emulsion Tester AET100

Agrochemical Emulsion Tester AET100

The Agrochemical Emulsion Tester AET100 is designed to test the stability of water-based agrochemical emulsions which are widely used in pest control. Easily surpassing current techniques, the turbidity fluctuation analysis technique can quickly and precisely characterise emulsions to bring significant benefits to testing in the development of new formulations and the quality of manufactured emulsifiable concentrate products.

The AET100 has been developed in collaboration with Zeneca Agrochemicals and has successfully completed UK trials involving major manufacturers.


Testing Emulsions

Good stable emulsions are characterised by very small droplets and by a fairly narrow distribution of droplet sizes, although the actual values will vary between different formulations. These properties should not change over periods of many hours or even longer.

The AET100 is designed to detect the size of droplets in oil-water emulsions using a simple procedure. A prepared emulsion sample is siphoned through a capillary sample tube mounted in a flow cell on the front of the instrument. After 1-2 minutes a reading related to the stability (R) is displayed: further tests can be conducted after flushing the sample tube clean.

Results achieved show good correlation with the traditional CIPAC test, but with significantly greater sensitivity and taking only a few minutes.

An optional validation unit is available to confirm the test performance of the instrument and for compliance with ISO9000 quality assurance procedures.


Operating Principle

Response of R to good and poor emulsions

The AET100 uses an optical system developed at University College London that responds to the changes in the size of droplets flowing through the measurement cell, as indicated in the diagram.

A laser light beam illuminates the flowing emulsion and the transmitted light is monitored by a sensitive detector, which produces a signal comprising two components:

In unstable emulsions, droplets become larger in size and smaller in number, which leads to an increase in the AC component. A measure of the emulsion stability is derived from the ratio of the AC and DC components, which is displayed as a reading R. The higher this reading, the larger the droplets and the lower the emulsion stability. Conversely, a small R value indicates a stable emulsion.


Applications

The stability of emulsions must be tested to ensure their efficient use and proper application. A rapid assessment of the emulsion stability using the AET100 will help to prevent breakdown of emulsions and the formation of oils and creams.

In developing a formulation

A plot of results using different emulsifier ratios enable the optimum formulation to be selected.

Selecting the optimum formulation

For routine quality monitoring

Frequent measurements quickly identify potential stability problems and corrective action can be taken during manufacture.


Technical Data

The AET100 is supplied with 6 capillary sample tubes, 2mm silicone tubing, Hoffman clip to adjust flow rate, 250ml beaker tubing clamp assembly and a suitable power lead.

Light source:
Laser, 670nm, output 0.95mW.
Sample flow rate:
10ml/min ±10%.
Measurement accuracy:
±5%.
Display:
Readings (R) and DC volts.
Overload indicators:
Large particles and turbidity.
Power supply:
110/120V or 220/240V AC 50/60Hz.
Size:
320mm (w) x 200mm (d) x 165mm (h).
Weight:
4.0kg.

CE Marked


Operating Manual

The manuals on this website are provided as PDF files, please see the Downloads page for links to reader software.

Agrochemical Emulsion Tester AET100 Manual


Ordering Information

Order Code Description
AET001A Agrochemical Emulsion Tester AET100 & validation unit
AET002A Agrochemical Emulsion Tester AET100
AET003A Validation unit for AET100
AET170A Glass sample tube
AET172A Silicone tubing (2 mm bore, 1 mm wall)

Rank Brothers Ltd reserves the right to change specifications in the light of continuing development.